Recall Z-1227-2018 ยท HSB Rod, Fixation, Intramedullary And Accessories

Zimmer Biomet, Inc.

ACE Trochanteric Nail System STER TROC NAIL 125 DEG X 9MM STER TROC NAIL 130 DEG X 9MM STER TROC NAIL 135 DEG X 9MM The Trochanteric Nail System is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures, intertrochanteric fractures, high subtrochanteric fractures and combinations of these fractures. The Trochanteric Long Nail System is additional

DateMar 14, 2018 (initiated Dec 20, 2017)
StatusTerminated
ReasonZimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system.
Root causeDevice Design
ActionOn December 22, 2017 an URGENT MEDICAL DEVICE RECALL letter was issued to customers with a complete list of products subject to recall and requesting that the customer review the notification and notify all affected personnel of the recall. If there is affected product at the facility it is requested that these units be quarantined and returned to a sales representative for return to the firm. Questions or concerns can be directed to 574-371-3071.
Quantity3454
DistributionUSA (nationwide)
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K010780TROCHANTERIC NAIL
DePuy Orthopaedics, Inc., Jun 12, 2001
K013563LONG TROCHANTERIC NAIL SYSTEM
DePuy Orthopaedics, Inc., Dec 27, 2001
K042325TROCHANTERIC NAIL SYSTEM
DePuy Orthopaedics, Inc., Sep 8, 2004
K111663TRAUMA INTERNAL FIXATION SYSTEMS
DePuy Orthopaedics, Inc., Mar 12, 2012