Recall Z-1226-2018 ยท HSB Rod, Fixation, Intramedullary And Accessories
Zimmer Biomet, Inc.
ACE Trochanteric Nail System STER TROC END CAP STER TROC END CAP 5MM STER TROC END CAP 10MM STER TROC END CAP 15MM STER TROC END CAP 20MM STER TROC END CAP 25MM The Trochanteric Nail System is intended to treat stable and unstable proximal fractures of the femur including pertrochanteric fractures, intertrochanteric fractures, high subtrochanteric fractures and combinations of these fractures. Th
| Date | Mar 14, 2018 (initiated Dec 20, 2017) |
| Status | Terminated |
| Reason | Zimmer Biomet is conducting a medical device recall for the ACE Trochanteric Nail (ATN) system after a Device Performance Review (DPR) failed clinical analysis for survivorship pursuant to the acceptance criteria for the DPR protocol for the ATN system. |
| Root cause | Device Design |
| Action | On December 22, 2017 an URGENT MEDICAL DEVICE RECALL letter was issued to customers with a complete list of products subject to recall and requesting that the customer review the notification and notify all affected personnel of the recall. If there is affected product at the facility it is requested that these units be quarantined and returned to a sales representative for return to the firm. Questions or concerns can be directed to 574-371-3071. |
| Quantity | 3454 |
| Distribution | USA (nationwide) |
| FDA record | Open on fda.gov |