ClearanceBrief
Recall Z-1196-2018 ยท HSB Rod, Fixation, Intramedullary And Accessories

Zimmer Biomet, Inc.

AIM Tibial Nails intramedullary fixation rod - STERILE; Item numbers: 800508255 NAIL TIB DYN 08X25.5 - STERILE 800508270 NAIL TIB DYN 08X27.0 - STERILE 800508285 NAIL TIB DYN 08X28.5 - STERILE 800508300 NAIL TIB DYN 08X30.0 - STERILE 800508315 NAIL TIB DYN 08X31.5 - STERILE 800508330 NAIL TIB DYN 08X33.0 - STERILE 800508345 NAIL TIB DYN 08X34.5 - STERILE 800508360 NAIL TIB DYN 08X36.0 - STERILE 80

DateMar 14, 2018 (initiated Feb 6, 2018)
StatusTerminated
ReasonThe AIM Tibial Nails failed a clinical analysis where the surgical revision rate exceeded the acceptance criteria in the Device Performance Review protocol for the implant system.
Root causeUnder Investigation by firm
ActionZimmer Biomet sent an Urgent Medical Device Recall letter dated February 06, 2018 to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to: 1. Review this notification and notify affected personnel of the contents. 2. Quarantine all affected product. Your Zimmer Biomet sales representative will remove it from your facility. 3. Complete the Certificate of Acknowledgement and send to CorporateQuality.PostMarket@zimmerbiomet.com. DISTRIBUTOR: 1. Review this notification and notify affected team members of the contents.2. Qu
QuantityNot known
DistributionWorldwide and US of Indiana and country of the Republic of Korea.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K972183DEPUY ACE A.I.M. TITANIUM TIBIAL NAIL
Ace Medical Co., Aug 26, 1997