Recall Z-1208-2018 ยท CBK Ventilator, Continuous, Facility Use

Covidien LLC

Rechargeable Li-ion battery used with Puritan Bennett TM 980 Series Ventilator System including Pediatric-Adult Ventilator, Neonatal Ventilator, Universal Ventilator; Battery Product Number: 10086042; UDI: 20884521700861 The rechargeable lithium-ion batteries are intended to provide a backup power source to the ventilator in case AC power is lost.

DateNov 21, 2017
StatusTerminated
ReasonRechargeable lithium-ion batteries with incorrect firmware that are used in certain Puritan Bennett 980 ventilators may not fully charge after installation.
Root causeProcess control
ActionConsignees are instructed to immediately notify all care environments in which the PB980 ventilators are used about the notification. If the facility has distributed PB980 ventilators to other persons or facilities, they are asked to please promptly forward a copy of the letter to those recipients. The consignee is asked to complete an attached form and return it as directed to confirm receipt and understanding of the notification.The firm developed a tool to inspect and identify the firmware on the batteries. Medtronic service engineers will inspect batteries at consignee facilities during sc
Quantity15,885
DistributionWorldwide Distribution - USA (nationwide) Distribution to the states of : AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, WA, WI, WV, WY, and the District of Columbia. and to the cou
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K131252PURITAN BENNETT 980 SERIES VENTILATOR SYSTEM
Covidien, Feb 24, 2014