Recall Z-1218-2018 ยท FQO Table, Operating-Room, Ac-Powered

Deerfield Imaging, Inc.

IMRIS ORT300 Removable Operating Room Table, 50-60 Hz Version is intended for use during diagnostic examinations or surgical procedures to support and position a patient.

DateFeb 23, 2018
StatusTerminated
ReasonThe rotational locking mechanisms on the ORT 200/300 tables fail to operate such that it cannot change states between locked and unlocked.
Root causeUnder Investigation by firm
ActionThe firm disseminated their notices on 02/23/2018 by mail and followed with an email on 02/26/2018. The notice stated: "It has recently come to our attention that the ORT200/300 rotational lock fails to operate such that it cannot change states between locked and unlocked. This may pose a risk of injury to the patient. To minimize the risk of injury, IMRJS is taking the following corrective actions: " ORT200 and 300 tables: all customer sites with these tables shall have the rotational lock operation verified and adjusted, if necessary. IMRIS has identified that your facility has one or mo
Quantity13 units
DistributionUS, Canada, China, Sweden, Germany, Japan, Qatar, France, Audtralia
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.