ClearanceBrief
Recall Z-1241-2018 ยท PHX Shoulder Prosthesis, Reverse Configuration

Encore Medical, Lp

AltiVate Reverse INSTRUMENTATION, HUMERAL SOCKET SHELL TRIAL, STANDARD, REF 804-06-152, Qty: 01, Encore Medical, LP Orthopedic surgical instrument. Product Usage: The humeral socket shell trial is used to replicate/trial the shell geometry of the AltiVate Reverse Standard Shell Humeral Stem implants. It assembles to the humeral broaches by a screw in order to replicate/trial the entire humeral s

DateMar 5, 2018 (initiated Jan 30, 2018)
StatusTerminated
ReasonBroken or missing screws and retaining clips. This may be due to screw tolerance and material strength of the trial.
Root causeDevice Design
ActionThe firm disseminated their field safety notices on 01/30/2018 by email. The letter identified the affected product, problem and actions to be taken. The notices requested the trails be returned for replacement.
Quantity387 units
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K141990ENCORE HUMERAL SHOULDER STEM
Encore Medical L.P., Nov 7, 2014