Recall Z-1927-2018 ยท FSY Light, Surgical, Ceiling Mounted

GETINGE US SALES LLC

ALM XTEN Surgical Light intended to provide visible illumination for the surgical field or for the examination of the patient. Part Numbers: ARD567801093, ARD567801094

DateMar 5, 2018 (initiated Nov 16, 2017)
StatusTerminated
ReasonGetinge has received complaints concerning the X TEN Surgical Lights whose spring arms have broken, causing light heads to separate from the assembly. The issue relates to cracking of the spring arms at the weld seams that has the potential to develop into breaks with heavy use and aging of the equipment.
Root causeUnknown/Undetermined by firm
ActionMaquet/Getinge notified customers via an Urgent Medical Device Recall Notice dated November 16, 2017.
Quantity667
DistributionU.S. and Gov. accts.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K040735ALM X'TEN (X10) SURGICAL LIGHT SYSTEM
Alm S.A., May 21, 2004