Recall Z-1216-2018 ยท HAW Neurological Stereotaxic Instrument
Stryker Instruments Div. of Stryker Corporation
The Stryker Thoracic Pedicle Feeler 6002-350-000 is a smart instrument for remote control of the navigation software and creation of a pathway into the pedicle. The instrument has three types of buttons to remotely control the software: the UP, DOWN and SELECT buttons. The Pedicle Feeler is laser marked in 10 mm intervals to help indicate the depth in which the feeler tip has been inserted into th
| Date | Oct 16, 2017 |
| Status | Terminated |
| Reason | Stryker Instruments and Stryker Leibinger GmbH & Co. KG are voluntarily recalling certain serial numbers Thoracic Pedicle Feelers that are used in conjunction with the Stryker Navigation System because the settings on the machine were higher than intended. This change led to increased energy directed onto the tip of the Thoracic Pedicle Feeler which caused material degradation where the laser marks exist. There is the potential for the tip of the Thoracic Pedicle Feeler to break. |
| Root cause | Software Manufacturing/Software Deployment |
| Action | US customers will be notified via FedEx overnight. US Sales Representatives will be notified via email. International entities will be notified via an internal notification system. International customers will be notified per local regulations/procedures. Customer Instructions: Actions to be taken by the Customer/User: 1. Immediately review this Recall Notification. 2. Immediately check all stock areas and/or operating room storage for affected equipment. Quarantine and discontinue use of any affected Thoracic Pedicle Feelers. 3. Complete the enclosed Business Reply Form (BRF) to confirm rec |
| Quantity | 28 |
| Distribution | International , US and US Govt. |
| FDA record | Open on fda.gov |