Recall Z-1260-2018 ยท KWS Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented
Encore Medical, Lp
Altivate Anatomic(TM) INSTRUMENTATION, PEGGED GLENOID TRIAL, SIZE 54mm, REF 804-07-540 Product Usage: The AltiVate Anatomic Shoulder System is a total shoulder system indicated as an anatomic shoulder joint replacement for patients suffering from pain and dysfunction.
| Date | Jan 29, 2018 |
| Status | Completed |
| Reason | It was discovered during a surgery that the design of the radius at the base of the pegs of the Anatomic Glenoid Trial (Part Numbers: 804-07-380, 804-07-420, 804-07-460, 804-07-500, and 804-07-540) is much larger on the trails than on the implants |
| Root cause | Device Design |
| Action | DJO Global sent an Urgent Field Safety Notice letter dated January 29, 2018, to affected customers. The letter identified the affected product, problem and actions to be taken. For questions call (512) 834-6255 |
| Quantity | 93 units |
| Distribution | US Nationwide Distribution |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K162024 | AltiVate Anatomic Shoulder System Encore Medical L.P., Nov 21, 2016 |