ClearanceBrief
Recall Z-1258-2018 ยท KWS Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented

Encore Medical, Lp

Altivate Anatomic(TM) INSTRUMENTATION, PEGGED GLENOID TRIAL, SIZE 46mm, REF 804-07-460 Product Usage: The AltiVate Anatomic Shoulder System is a total shoulder system indicated as an anatomic shoulder joint replacement for patients suffering from pain and dysfunction.

DateJan 29, 2018
StatusCompleted
ReasonIt was discovered during a surgery that the design of the radius at the base of the pegs of the Anatomic Glenoid Trial (Part Numbers: 804-07-380, 804-07-420, 804-07-460, 804-07-500, and 804-07-540) is much larger on the trails than on the implants
Root causeDevice Design
ActionDJO Global sent an Urgent Field Safety Notice letter dated January 29, 2018, to affected customers. The letter identified the affected product, problem and actions to be taken. For questions call (512) 834-6255
Quantity92 units
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K162024AltiVate Anatomic Shoulder System
Encore Medical L.P., Nov 21, 2016