ClearanceBrief
Recall Z-1146-2018 ยท KRA Catheter, Continuous Flush

Vascular Insights, LLC

ClariVein IC infusion catheter, Model Number: 65-018-E4S Product Usage: The ClariVein IC infusion catheter is indicated for infusion of physician-specified agents into the peripheral vasculature.

DateNov 9, 2017
StatusTerminated
ReasonCertain catheters could have a compromised catheter shaft due to a manufacturing defect which could result in some delamination of the catheter shaft including during introduction of the catheter into a vascular access introducer.
Root causeProcess control
ActionVascular Insights sent an Urgent Voluntary Medical Device Recall letter to their customers. The letter identified the affected product, problem and the actions to be taken. Customers were instructed to 1. Do not use or distribute any recalled product. 2. Immediately remove all recalled product from your inventory. 3. Segregate recalled product in a secure location and follow the instruction herein to return the recalled product and receive replacement product. 4. Immediately forward a copy of this recall notification to any/all sites to which you may have distributed the affected prod
Quantity928
DistributionU.S Nationwide in the states: AZ, CA, CT, FL, GA, IA, IL, IN, KS, KY, MA, MD, MO, NE, NJ, NY, OH, PA, SC, TN, TX, VA, WA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071468CLARIVEIN CATHETER
Vascular Insights, LLC, Mar 20, 2008