Recall Z-1285-2018 ยท DQX Wire, Guide, Catheter

Arrow International Inc

Radial Artery Catheterization Kit; Model: AK-04122

DateOct 26, 2017
StatusTerminated
ReasonThe povidone iodine prep pads included in certain kits and trays do not meet the requirements for a 36-month shelf life and should be labeled with a 24-month shelf life.
Root causeNonconforming Material/Component
ActionTeleflex/Arrow sent a letter dated October 26th, 2017. Customers are instructed to discontinue use and quarantine any products with the affected product codes and lot numbers. There is a Recall Acknowledgment Form that is requested to be completed that will allow the return and replacement of the affected product. Consignees without affected product on hand are also asked to return the form.
Quantity570
DistributionUS nationwide distribution, including Puerto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K810675ARROW RADIAL ARTERY CATHETERIZATION SET
Arrow Intl., Inc., Mar 27, 1981