Recall Z-1066-2018 ยท GJS Test, Time, Prothrombin

Diagnostica Stago, Inc.

Neoplastine Cl Plus @ (ref.00374) Product Usage: Manual or automated determination of the prothrombin time (PT).

DateOct 25, 2017
StatusTerminated
ReasonConfirmed defect of homogeneity where some reagent vials will give prolonged Prothrombin Time (decreased PT%).
Root causeDevice Design
ActionStago sent an Urgent Field Safety Notice letter dated October 25, 2017 to their customers. The letter identified the affected product, problem and actions to be taken: Customers were instructed to: If it is not already done, run a Quality Control test at every change of vial (if you have affected lots of product). Return to company, by fax or email, an acknowledgement form confirming that the recall letter was read by the receiving company. For questions contact the Stago Hotline (1-800-725-0607).
Quantity1636 Kits
DistributionNationally
FDA recordOpen on fda.gov