Recall Z-1296-2018 ยท DOF Lead, Atomic Absorption

Magellan Diagnostics, Inc.

LeadCare Plus Blood Lead Analyzer Part Number: 82-0001 Intended for the quantitative measurement of lead in a whole blood sample.

DateJan 9, 2018
StatusTerminated
ReasonThe incorrect placement of AA batteries into certain LeadCare Plus Blood Lead Analyzer may cause the batteries to overheat, and the chance that a user could burn their fingers if the batteries are removed while hot.
Root causeDevice Design
ActionAn Urgent Medical Device Correction letter, dated January 9, 2018, was sent to customers. The letter identified the affected device and provided a description of the issue. If two of the four AA batteries are inserted incorrectly, the batteries may overheat, causing a minor burn if the user opens the battery compartment and tries to remove the hot batteries. Customers are to follow the steps outlined in the letter, including placing the enclosed label on the upper back panel of the analyzer to remind users who are changing batteries of the importance of correct placement. Proper battery pl
Quantity79
DistributionWorldwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K142705LeadCare Plus Blood Lead Testing System
Magellan Diagnostics, Jul 7, 2015