Recall Z-1006-2018 ยท IYE Accelerator, Linear, Medical

Brainlab AG

ExacTrac versions 6.5 through 6.5 intended to be used to place patients at an accurately defined point within the treatment beam of a medical accelerator for stereotactic radiosurgery or radiotherapy procedures, in order to treat lesions, tumors and conditions anywhere in the body when radiation treatment is indicated. ExacTrac may also be used to monitor the patient position during the treatment.

DateFeb 14, 2018
StatusTerminated
ReasonThe usage of workflows that deviate from the recommended specifications in the User manual for ExacTrac Patient Positioning System (versions 6.0, 6.1, 6.2, 6.5) with Auxiliary Device Interface (ADI) and Varian Clinac or Varian TrueBeam, which may result in misinterpretation of beam authorization via ADI.
Root causeOther
ActionThe firm disseminated their notices by email on 02/14/2018. The notice requests the following actions: "User Corrective Action: Always follow the instructions below during use of ExacTrac Patient Positioning System (versions 6.0, 6.1, 6.2, 6.5) with Auxiliary Device Interface (ADI) and Varian Clinac or Varian TrueBeam: Do not use beam authorization by ExacTrac via ADI as a safety feature to ensure correct patient positioning before treatment. For any treatments performed with ExacTrac always verify that ExacTrac confirms the target position by displaying the OK icon. Do not treat the patie
Quantity587 systems
Distributionworldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K120789EXACTRAC X-RAY X-RAY
Brainlab AG, Aug 14, 2012