Recall Z-1149-2018 ยท CHI Lipase-Esterase, Enzymatic, Photometric, Lipase
Medica Corporation
Wash Kit (REF 10680) Lots 16279, 17045 and 17215 Medica corporation is marketing a wash solution Chemistry Wash1 , REF # 10680 for use with Medica s EasyRA clinical chemistry analyzer. The solution is used prior to a lipase (LIP) assay if an interferant assay (TRIG)is part of the work list. The major interferant for Lipase is the Triglycerides (TRIG) assay since one of the reagents contains lipas
| Date | Nov 21, 2017 |
| Status | Terminated |
| Reason | The firm has initiated a field recall of the Wash1 Kit REF 10680 due to reported inconsistent Lipase values when analyzing patient samples. The firm's internal investigation identified the underlying cause to be an ineffective wash solution resulting in an interference carryover from the triglyceride (TRIG) assay if they are on the same worklist. |
| Root cause | Under Investigation by firm |
| Action | Firm emailed all consignees and instructed them to remove/discard affected wedges (bottles) and contact Medica for immediate replacement 1 Immediately discontinue use of the Wash1 wedges (REF 10680) (Lot# listed above) and discard any remaining kits/wedges. 2. Please discuss possible concerns about lipase results achieved using these lots of Wash wedges with your laboratory manager and/or clinician in order to decide on appropriate measures. Report any concerns to Medica for evaluation and review. 3.Submit claims for replacement of affected product to your authorized Medica EasyRA deale |
| Quantity | 392 |
| Distribution | Worldwide and US Distribution. |
| FDA record | Open on fda.gov |