Recall Z-1072-2018 ยท MBI Fastener, Fixation, Nondegradable, Soft Tissue

Zimmer Biomet, Inc.

TunneLoc Tibial Fixation Device (Within an ACL Kit) Product Usage: To provide fixation of soft tissue grafts within the tibial tunnel during anterior cruciate ligament (ACL) and/or posterior cruciate ligament (PCL) reconstruction.

DateOct 11, 2017
StatusTerminated
ReasonGamma sterilization validation testing indicates the sterilization dose for the product may not be adequate.
Root causeVendor change control
ActionZimmer Biomet sent an URGENT MEDICAL DEVICE RECALL letter dated October 16, 2017 to affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to review inventory and quarantine affected product, which will be returned to Biomet through a sales representative. Questions or concerns can be directed to 574-371-3071
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K103145HAMMERLOC TM TIBIAL FIXATION DEVICE
Biomet Sports Medicine, Feb 9, 2011