Recall Z-1012-2018 ยท KWS Prosthesis, Shoulder, Semi-Constrained, Metal/Polymer Cemented

DePuy Orthopaedics, Inc.

The GLOBAL UNITE Platform Shoulder System is intended for cemented or uncemented total shoulder or hemi-shoulder replacement procedures, Product labeling: LEO-1100-30-000 REV D.pdf.pdf, LEO-1100-40-000 REV D.pdf.pdf, LEO-110030100_REV D.pdf.pdf, LEO-110030110_REV D.pdf.pdf, LEO-110030120_REV D.pdf.pdf

DateFeb 27, 2018 (initiated Nov 28, 2017)
StatusTerminated
ReasonThe screw in specific lots of the GLOBAL UNITE Anatomic Body and GLOBAL UNITE Fracture Body was inverted during assembly to the body, which will cause the humeral stem to sit proud and may cause surgical delays.
Root causeEmployee error
ActionDePuy Orthopaedics, Inc. is voluntarily recalling 11 lots of the GLOBAL UNITE Platform Shoulder System. The device is being recalled because the screw in specific lots of the GLOBAL UNITE Anatomic Body and GLOBAL UNITE Fracture Body was inverted during assembly to the body, which will cause the humeral stem to sit proud and may cause surgical delays. Once the screw is properly assembled, there is no additional risk to the patient. Further distribution or use of the affected lots is to cease immediately.DePuy Orthopaedics, Inc. is not recommending prophylactic revision. The company recommends t
Quantity85
DistributionDistribution Level: The device removal is extended to the companies US Distributors/Sales Offices and Medical Facilities that may have used, purchased, or received the affected devices. Approximately 53 units distributed to firm's international affiliates.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133834DEPUY GLOBAL UNITE SHOULDER SYSTEM
Depuy(Ireland), Apr 3, 2014