Recall Z-0828-2018 ยท FSY Light, Surgical, Ceiling Mounted

Stryker Communications

Berchtold Chromophare Ceiling Mounted Surgical Light System, Catalog #CH13000000. Product Usage: Chromophare surgical lights are medical lights for usage in hospital treatment rooms. They provide localized lighting focused on the patient s body for detecting and treating disease, injury, and disability. The surgical lights may only be used in rooms used for medical purposes that have been equippe

DateMar 1, 2018 (initiated Jan 29, 2018)
StatusTerminated
ReasonThe device could have a missing spanner nut and snap ring which could potentially result in the equipment falling.
Root causeEmployee error
ActionStryker sent an Urgent Medical Device Correction letters to affected customers. The letter identified the affected product, problem and actions to be taken. The customer is informed a firm representative will be conducting a visual inspection of the equipment. No response form was enclosed. For questions email CommunicationsRA1@stryker.com
Quantity1,162 devices
DistributionWorldwide Distribution - US Nationwide Distribution and in the countries of Canada.