Recall Z-0812-2018 ยท CBK Ventilator, Continuous, Facility Use

ZOLL Medical Corporation

731 Series Ventilators running software version 05.20.00 The devices in the ZOLL ventilator are indicated for use in the management of infant through adult patients weighing ;:: 5 kg with acute or chronic respiratory failure or during resuscitation by providing continuous positive-pressure ventilation. They are appropriate for use in hospitals, outside the hospital, during transport and in sever

DateJun 30, 2017
StatusTerminated
ReasonA software anomaly in the 731 software version 05/20/00, was identified, which can lead to a user inadvertently changing device settings.
Root causeUnder Investigation by firm
ActionOn June 30, 2017 an urgent recall letter was mailed to customers with the following instructions: Customers who have affected devices should immediately take the following steps: (1) Alert all users of 731 Ventilator models to this problem. (2) Direct users to always verify device parameters after selecting the device Confirm button. (3) Contact ZOLL's Technical Service Department or your local ZOLL Service Provider to schedule a software update for affected devices Our 24/7 technical support numbers 1(800) 348-9011 or +l (978) 421- 9460 are available to assist users with any aspect of this no
Quantity46
DistributionOutside US
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071526UNI-VENT 731 SERIES VENTILATORS, MODEL 731, 731EMV
Impact Instrumentation, Inc., Dec 10, 2008
K091238UNI-VENT-731 SERIES VENTILATORS, MODEL 731EMV+
Impact Instrumentation, Inc., Aug 13, 2009
K103318UNI-VENT 731 SERIES MODEL EMV+ PORTABLE CRITICAL CARE VENTILATOR
Impact Instrumentation, Inc., Apr 7, 2011
K111473UNI-VENT (R) MODEL 731 SERIES PORTABLE CRITICAL CARE VENTILATOR
Impact Instrumentation, Inc., Feb 16, 2012