ClearanceBrief
Recall Z-0993-2018 ยท LZO Prosthesis, Hip, Semi-Constrained, Metal/Ceramic/Polymer, Cemented Or Non-Porous, Uncemented

Zimmer Biomet, Inc.

JUGGERKNOT 2.9MM 2 #2 MB Product Usage: For use in hip arthroplasty

DateJun 30, 2017
StatusTerminated
ReasonFor soft tissue fixation
Root causePackaging process control
ActionOn June 2, 2017 an URGENT Medical Recall Letter was issued to customers regarding affected product(s). The letter requests customers to review the content and notify affected personnel of the recall. Assist the sales representative in quarantining affected product, which will be removed and returned to recalling firm and then replaced. All customers who may have received product from distributors are to be notified. Questions or concerns can be directed to 1-574-372-4847.
Quantity50
DistributionUS in the state of Texas
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K131684BIOLOX(R) DELTA CERAMIC HEADS
Biomet, Inc., Oct 10, 2013