Recall Z-0829-2018 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Fresenius Medical Care Renal Therapies Group, LLC

2008T, Hemodialysis Delivery System Product Indicated for acute and chronic dialysis therapy.

DateFeb 20, 2018 (initiated Oct 5, 2017)
StatusTerminated
ReasonWhile reviewing documentation for the next software release of the 2008T, an R&D technician identified that setting the UF goal to '0' introduces a discrepancy between the UF rate displayed and the actual UF pump rate. The software anomaly is also applicable to the 2008T Machines that contained the 2.63 Bug Fix per DCAF 17-088. 2008T Upgrade kits that contained the 2.63 Bug Fix are also affected. Additionally, SW version 2.64 (in design freeze) is also impacted by this anomaly.
Root causeSoftware Design Change
ActionFresenius Medical Care sent an Urgent Medical Device Correction letter dated April 24, 2017 to affected customers. the letter identified the affected product, problem and actions to be taken. For questions contact Technical Services at 1-800-227-2572.
Quantity15 machines
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K080964FRESENIUS MODEL 2008T HEMODIALYSIS MACHINE
Fresenius Medical Care North America, Jan 14, 2009
K113427FRESENIUS 2008T HEMODIALYSIS MACHINE
Fresenius Medical Care North America, Dec 21, 2011
K120017FRESENIUS 2008 HEMODIAYSIS MACHINE WITH BIBAG SYSTEM
Fresenius Medical Care North America, Feb 2, 2012
K120505FRESENIUS 2008T HEMODIALYSIS MACHINE
Fresenius Medical Care North America, Design Cente, Mar 6, 2012
K121341FRESENIUS 2008T HEMODIALYSIS MACHINE WITH BIBAG SYSTEM
Fresenius Medical Care North America, Dec 6, 2012