Recall Z-1010-2018 ยท KPO Dialysate Concentrate For Hemodialysis (Liquid Or Powder)

Fresenius Medical Care Renal Therapies Group, LLC

NaturaLyte Liquid Acid Concentrate 2.0 mEq/L, Model # 08-2251-0 3.43 liter bottle This concentrate is formulated to be used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.

DateSep 29, 2017
StatusTerminated
ReasonThe label specifies the potassium concentration is 2.0 mEq/L. Some of the product in the affected lot was found to have a potassium concentration of 2.25 mEq/L.
Root causeRelease of Material/Component prior to receiving test results
ActionThe firm, Fresenius Medical Care, sent an "URGENT - MEDICAL DEVICE RECALL" letter dated 10/4/2017 to its Consignees. The letter described the product, problem and actions to be taken. The Consignees are instructed to: " Immediately examine your stock to determine whether you have any Naturalyte Liquid Acid Concentrate from the lot listed. " If any product of this lot is found, discontinue use immediately. " Place all units in a secure, segregated area. " Contact your FMCRTG, LLC Customer Service Team at 1-800-323-5188 for instructions on how to return the recalled product. " Promptly fill out
Quantity598 cases
DistributionUS Distribution to states of: CA, FL, KS, LA, and TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K070177FRESENIUS NATURALYTE LIQUID ACID CONCENTRATE, 9000, 6000 AND 4000 SERIES
Fresenius Medical Care North America, Mar 29, 2007