Recall Z-0900-2018 ยท DWJ System, Thermal Regulating

Medtronic Perfusion Systems

BIO CAL(R) Temperature Controller, Models: (a) 370, (b) 370I Product Usage: The BIO CAL(R) (unit) is a thermal transfer module designed to control patient blood temperature during cardiopulmonary bypass procedures when heating or cooling of the blood is desired.

DateFeb 8, 2018
StatusTerminated
ReasonMedtronic has been unable to develop a cleaning protocol to satisfy current industry concerns and expectations. As a result, an updated cleaning protocol will not be developed by Medtronic and it has been determined that the best course of action is to request BIO CAL users to discontinue use and dispose of BIO CAL devices.
Root causeDevice Design
ActionMedtronic sent an Urgent Medical Device Recall dated February 2018 via UPS 2-Day.to their customers. The letter identified the affected product, problem and actions to be taken. Medtronic recommended that the affected device no longer be used for clinical procedures. Customers are asked to dispose of any devices in their possession per their normal equipment obsolescence procedures. Customer are also instructed to not resell any devices for clinical use. For questions contact your Medtronic Sales Representative.
Quantity1980 units
DistributionWorldwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K894980BIO-CAL 370 HEATER/COOLER
Bio Medicus, Inc., Nov 2, 1989