Recall Z-0938-2018 ยท LHN System, Radiation Therapy, Charged-Particle, Medical

Ion Beam Applications S.A.

Proteus 235, Proton Therapy System a medical device designed to produce and delivers a proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation.

DateNov 17, 2017
StatusTerminated
ReasonThere is a risk of misalignment of the patient when using treatment plans with multiple isocenters.
Root causeDevice Design
ActionOn November 17, 2017 an Urgent Field Safety Notice was issued to customers informing users that when creating or importing a treatment plan with multiple isocenters, the user has to enter Patient Positioning System positions manually.
Quantity37
DistributionNationally
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K101508MODIFICATION TO IBA PROTON THERAPY SYSTEM-PROTEUS 235
Ion Beam Applications S.A., Aug 6, 2010