Recall Z-1653-2018 · CGS Nad Reduction/Nadh Oxidation, Cpk Or Isoenzymes
Siemens Healthcare Diagnostics, Inc.
Dimension Creatine Kinase MB (MBI) assay (Product Number: DF32/10464510) (Mfr: Siemens Healthcare Diagnostics, Newark, DE)
| Date | Jun 29, 2017 |
| Status | Terminated |
| Reason | Various assays used to measure different chemical levels in blood, urine, or cerebrospinal fluid may interfere with the drugs sulfasalazine and/or sulfapyridine causing falsely elevated or falsely depressed results. |
| Root cause | Device Design |
| Action | Letters were sent to consignees with the following instructions: " Please review this letter with your Medical Director. " Venipuncture should occur before drug administration of Sulfasalazine or Sulfapyridine as indicated above under Reason For Correction. Baseline assay values before administration of Sulfasalazine or Sulfapyridine therapy would not be affected. " Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. " If you have received any complaints of illness or adverse events associated with the products listed in Table 1 or 2, immed |
| Quantity | 30,446 |
| Distribution | Distributed in all 50 U.S. states, the District of Columbia, Guam, and Puerto Rico. Distributed in 108 foreign countries: Afghanistan Bolivia Albania¿¿¿¿¿¿¿¿ Algeria¿¿¿¿¿¿¿¿ Angola ARGENTINA Armenia AUSTRALIA Austria¿¿¿¿¿¿¿¿ Azerbaijan Bahamas Bahrain¿¿¿¿¿¿¿¿ Bangladesh¿¿¿¿¿ Belarus¿¿¿¿¿¿¿¿ Belgiu |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K081731 | DIMENSION CKI FLEX REAGENT CARTRIDGE, DIMENSION MBI FLEX REAGENT CARTRIDGE, AND CKI/MBI CALIBRATOR, MODELS DF38, DF32 Siemens Healthcare Diagnostics, Aug 29, 2008 |