Recall Z-1373-2018 ยท LXH Orthopedic Manual Surgical Instrument

TriMed Inc.

Hex Driver Part Number HXDRIVER-2.0 AO; Hex Driver Part Number HXDRIVR-2.5 AO; Hex Driver Part Number HXDRIVR-2.5CL AO; Hex Driver Part Number DVHX-2.5/130C AO; Hex Driver Part Number HXDRIVR-3.0 AO; Hex Driver Part Number HXDRIVER-3.5AO

DateJan 11, 2018
StatusTerminated
ReasonFollowing the completion of sterilization cycle(s) of the impacted Hex Driver, the epoxy color band on the impacted Hex Driver can migrate down the shaft and/or potentially onto the sterilization tray or caddy.
Root causeProcess control
ActionRecall notification letters were sent to affected consignees on January 16, 2018. The consignee letter requests that the consignee return an initial Acknowledgement & Receipt Form to TriMed by January 23, 2018.
Quantity1525 devices
DistributionUS, China, Latin America
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.