Recall Z-1309-2018 ยท EZD Catheter, Straight

C.R. Bard, Inc.

Personal Intermittent Catheters- 16" and 10" lengths - 30 catheters per box, each catheter is packaged in an individual pouch, Item Numbers: 50612, 50614, 50616, 50618, 52306, 52308, 52310, 52606, 52606G, 52608, 52608G, 52610, 52610G, 53314, 53320, 53514G, 53610, 53610G, 53612G, 53614G, 53616G, 53618G, 62308, 62310, 62606P, 602608, 62608P, 62610, 62610P, 63310, 63316, 63610, 63610M, 63612, 63612

DateApr 4, 2018 (initiated Sep 14, 2017)
StatusTerminated
ReasonSeveral instances of a catheter caught in the sterile barrier seal and in some instances cut catheters were observed.
Root causeProcess control
ActionA notification letter was sent to four hundred and fourteen (414) consignees via Federal Express on September 15, 2017. The letter explained the problem and requested return of the product. Distributors were directed to cease distribution of the recalled products and notify their customers of the recall.
Quantity10,949,013 units in total
DistributionUS Nationwide in the states of AK, AL, AR, AZ, CA, FL, GA, HI, IA, ID, KS, KY, MD, ME, MI, MN, MO, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA, WI, WV, including Puerto Rico