ClearanceBrief
Recall Z-0965-2018 ยท DQY Catheter, Percutaneous

Cook Inc.

Left Atrial Catheter Set

DateOct 13, 2017
StatusTerminated
ReasonLabel does not state that the product is heparin-coated.
Root causeLabeling Change Control
ActionOn October 31, 2017 an URGENT MEDICAL DEVICE CORRECTION was issued to customers requesting the following, including a corrective action: Step 1: Determine if your product is included in the affected lot ranges by comparing your lot number to the listing provided. " If your product is not affected, complete the Acknowledgement and Receipt Form. " If your product is affected, proceed to Step 2. Step 2: Remove a Heparin Coated label from the label sheet provided and apply to the top right corner of your affected unit. o If you need additional labels, please contact Stericycle at 855.215.4967. If
Quantity56
DistributionNationally
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.