ClearanceBrief
Recall Z-0978-2018 ยท MKJ Automated External Defibrillators (Non-Wearable)

Philips Electronics North America Corporation

HeartStart XL+ Defibrillator/Monitor The HeartStart XL+ is intended for use in a hospital setting by qualified medical personnel trained in the operation of the device and qualified by training in basic life support, advanced life support, or defibrillation. It must be used by or on the order of a physician. When operating as a semi-automatic external defibrillator in AED Mode, the HeartStart XL+

DateFeb 7, 2018
StatusTerminated
ReasonUpdate XL device software to version A.03. This includes enhancements to the Operational Checks, event logs, and troubleshooting messages that provide as complete of information as needed to users on device readiness.
Root causeLabeling mix-ups
ActionPhilips is initiating a correction to affected devices. A software upgrade is available free of charge for all units affected by the issue. A Philips Healthcare representative will contact customers with affected devices to arrange for installation of the upgrades. Customers may continue to use XL+ prior to receiving the software upgrade. For further questions, please call (978) 659-3000.
Quantity4315
DistributionForeign Only - China
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K110825PHILIPS HEARTSTART XL+, PHILIPS HEARTSTART EXTERNAL PADDLES, PHILIPS MULTIFUNCTION ELECTRODES
Philips Medical Systems, Feb 16, 2012