Recall Z-0915-2018 ยท DIF Drug Mixture Control Materials

Ortho-Clinical Diagnostics

VITROS Chemistry Products DAT Performance Verifier V in vitro diagnostic

DateSep 22, 2017
StatusTerminated
ReasonThis product is not meeting the current storage labeling for unopened VITROS DAT Performance Verifiers causing customers to obtain lower than expected VITROS BENZ results when using this control.
Root causeUnder Investigation by firm
ActionOn September 18, 2017 a letter was sent to customers who received the BENZ reagent within the past 18 months of affected lots. The letter stated that customers may continue use of the product if QC results using VITROS DAT Performance Verifiers are within established ranges for each assay. If QC BENZ results are below established ranges, the customer should discontinue use of the product. The letter requests customers post the notification by each system that processes the VITROS BENZ Reagent. If the product was transferred outside of the facility, to notify those customers as well. Affected
Quantity102
DistributionWorldwide distribution: US Nationwide (including Puerto Rico), Bermuda, Australia, Brazil, Canada, Chile, Colombia, India, Japan, Mexico, Panama, Singapore, Venezuela, England, France, Germany, Italy, Spain, Poland, and Russia.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K062077VITROS CHEMISTRY PRODUCTS AMPH REAGENT; CALIBRATOR KIT 26; FS CALIBRATOR 1; DAT PERFORMANCE VERIFIERS
Ortho-Clinical Diagnostics, Inc., Nov 22, 2006
K062165VITROS CHEMISTRY PRODUCTS BARB REAGENT; CALIBRATOR KIT 26; FS CALIBRATOR 1; DAT PERFORMANCE VERIFIERS
Ortho-Clinical Diagnostics, Jan 5, 2007