ClearanceBrief
Recall Z-0707-2018 ยท IZI System, X-Ray, Angiographic

Siemens Medical Solutions USA, Inc

Artis Zee and Zeego, Model Numbers: 10094135, 10094137, 10094141, 10094142, 10094143, 10280959, 10848283 & 10848354

DateJun 6, 2017
StatusTerminated
ReasonPotential for liquid entry
Root causeDevice Design
ActionOn June 6, 2017, Siemens Healthineers distributed URGENT MEDICAL DEVICE CORRECTION notices to their customers via USPS Certified Mail. Siemens conducted a corrective field action in June 2015 to protect the Artis table against liquid entry at critical locations. However, even with this corrective action, if there are larger amounts of liquid on the table surface, they can still enter the table. Siemens Medical Solutions USA, Inc. will contact customers to arrange a date to perform the corrective action. Please feel free to contact the service organization for an earlier appointment by callin
DistributionU.S. Nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K073290ARTIS ZEE AND ZEEGO
Siemens Medical Solutions USA, Inc., Feb 11, 2008