Recall Z-0707-2018 ยท IZI System, X-Ray, Angiographic
Siemens Medical Solutions USA, Inc
Artis Zee and Zeego, Model Numbers: 10094135, 10094137, 10094141, 10094142, 10094143, 10280959, 10848283 & 10848354
| Date | Jun 6, 2017 |
| Status | Terminated |
| Reason | Potential for liquid entry |
| Root cause | Device Design |
| Action | On June 6, 2017, Siemens Healthineers distributed URGENT MEDICAL DEVICE CORRECTION notices to their customers via USPS Certified Mail. Siemens conducted a corrective field action in June 2015 to protect the Artis table against liquid entry at critical locations. However, even with this corrective action, if there are larger amounts of liquid on the table surface, they can still enter the table. Siemens Medical Solutions USA, Inc. will contact customers to arrange a date to perform the corrective action. Please feel free to contact the service organization for an earlier appointment by callin |
| Distribution | U.S. Nationwide |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K073290 | ARTIS ZEE AND ZEEGO Siemens Medical Solutions USA, Inc., Feb 11, 2008 |