Recall Z-0842-2018 ยท MBQ Peripheral Electromagnetic Field (Pemf) To Aid Wound Healing

Toshiba American Medical Systems Inc

INFX-8000F Fluoroscopic X-Ray Systems

DateDec 29, 2017
StatusTerminated
ReasonDuring an examination a device error occurred and the ceiling drive operation was disabled. It was found that the fastening bolts of the driving portion of the ceiling drive loosened cause the error.
Root causeProcess control
ActionA recall notification letter was sent to affected consignees on 12/29/2017. The following information was included in the notification letter: Should this problem occur prior to the above corrective action and in the middle of a clinical case, you will be able continue the case tentatively. The x-ray emissions will be still available and the inhibited ceiling support device can be overridden by pressing the override switch on the left side of the operation console. If any abnormalities are found during use, please stop using your system and contact your Toshiba service representative. Please s
Quantity17 units
DistributionUnited States
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081624INFX-8000F (INFINIX CSI AND INFINIX DPI)
Toshiba America Medical Systems, In.C, Jun 23, 2008