Recall Z-0926-2018 ยท JQP Calculator/Data Processing Module, For Clinical Use

Abbott Point Of Care Inc.

i-STAT DE handheld data processing module for clinical use, Software Version 2.8, List Number:08K46-01115200 (UDI: 00054749001255)

DateSep 12, 2017
StatusTerminated
ReasonIssues resulting from upgrade to software version 2.8: (1) Location, operator, stored patient lists will not update, and (2) Customized Reference Ranges, Action Ranges, and Custom Reportable Ranges are reset to factory default values. No erroneous results are generated as a result of this issue.
Root causeDevice Design
ActionThe following actions are recommended if you have already upgraded to version 2.8: 1. If you are using Location, Operator, or Patient lists. a. Delete the following files: i. C:\Istat32\LocInstData\LocInstData.txt ii. C:\Istat32\OpData\OpData.txt iii. C:\Istat32\PatData\PatData.txt iv. C:\Istat32\Log\iSTATDMI Errors.txt v. All files named C:\Istat32\Log\DeCustLogXX.txt; where XX is any two digits. b. Stop and restart the DECOMSRV service, which in turn restarts the DEDPSRV service. 2. If you are using customized Reference Ranges, Action Ranges, and Custom Reportable Ranges: a. Open DE Customiz
Quantity894
DistributionDistributed in 48 States: AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, the District of Columbia, Australia, Italy, Sweden, and Canada.