Recall Z-0704-2018 ยท MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)

Philips Electronics North America Corporation

digital x-ray detector ProGrade R1 - solid state X ray imager (flat panel/digital imager) As a part of a radiographic system, the Philips ProGrade is intended to acquire, process, store, display, and export digital radiographic images. The Philips ProGrade is suitable for all routine radiographic examinations, including specialist area like intensive care, trauma, or pediatric work, excluding ma

DateDec 11, 2017
StatusTerminated
ReasonIt the WiFi connection between the SkyPlate detector and HP transfer point is weak, an image may fail to transfer from the SkyPlate detector to the system. The image remains in the memory of the detector, but cannot be transferred wirelessly or by use of the backup cable. To continue, the operator can reset the SkyPlate detector by removing its batteries, but the acquired image is lost and a re-take is necessary.
Root causeUnder Investigation by firm
ActionPhilips sent an Urgent- Field Safety Notice date December 11, 2017. Philips plans to release a new HP firmware to the installed base. In addition the existing Access Point Hardware in the field needs to be patched. A Philips Service Engineer will contact the affected Customers when the Field Action Kit is available to be implemented. For further questions, please call (978) 659-7275.
Quantity157
DistributionWorldwide Distribution - USA (nationwide) Distribution, Internationally to Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K141736PHILIPS PROGRADE; PHILIPS ELEVA WORKSPOT FOR DIGITAL DIAGNOST
Philips Medical Systems Dmc GmbH, Jul 25, 2014