Recall Z-0823-2018 ยท MUU System, Suction, Lipoplasty

LifeCell Corporation, A Wholly Owned Subsidiary of Allergan PLC

REVOLVE ADVANCED ADIPOSE SYSTEM, for aspiration, harvesting, filtering, and transferring of autologous adipose tissue for aesthetic body contouring.

DateAug 8, 2017
StatusTerminated
ReasonPresence of bacterial endotoxins levels above the acceptable limit
Root causeMaterial/Component Contamination
ActionThe firm's initial notice to customers (dated May 25, 2017) advising customers that Allergan has placed a hold on shipping the device until verification is complete. Within this notice, customers were advised to continue use of the product. On June 14, 2017, a second notice was issued to customers informing customers that product shipments will resume by August 1, 2017 and that manufacturing process improvements will be implemented. On August 8, 2017, Allergen distributed URGENT MEDICAL DEVICE RECALL notices and Business Reply Forms (dated August 2, 2017) to US customers advising them to t
Quantity25,798 units
DistributionUS Nationwide, Canada, Australia, Chile, Brazil, Israel, Japan New Zealand & Singapore