ClearanceBrief
Recall Z-0615-2018 ยท FAJ Cystoscope And Accessories, Flexible/Rigid

Pentax of America Inc

Video Cystoscope models ECY-1570 and ECY-1570K

DateJul 13, 2017
StatusTerminated
ReasonThe video cytoscopes lack 510(k) premarket notification clearance.
Root causeNo Marketing Application
ActionCustomer Instructions: Customers should inspect their stock of PENTAX Medical ECY-1570 and ECY-1570K video cystoscopes, quarantine them, and discontinue use of these products. If you have sold and distributed these products, please provide a copy of the notice and response form to the new owners. Enclosed with this letter is a customer response form. The form identifies the affected endoscopes (model and serial numbers) which have been sold to your facility. The attached form should be completed and returned to PENTAX Medical. Contact Information: If you have any questions regarding this actio
Quantity24
DistributionDistributed in 8 states: CA, IL, MD, MI, ND, NH, NY, SD, and Guam.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.