Recall Z-0703-2018 ยท IXR Table, Radiographic, Tilting
Oakworks Inc
Foot Controls impacted for the following devices: OAKWORKS Medical CFPM 300 Imaging Table OAKWORKS Medical CFPM 301 Imaging Table OAKWORKS Medical CFPM 400 Imaging Table OAKWORKS Medical CFPM 401 Imaging Table OAKWORKS Medical CFPMB 301 Bariatric Imaging Table OAKWORKS Medical CFUR 301 Urology Table OAKWORKS Medical CFUR 401 Urology Table OAKWORKS Medical CFLU 401 Lithotripsy/Urology Table
| Date | Feb 23, 2018 (initiated Jul 31, 2017) |
| Status | Terminated |
| Reason | The foot control used on the OAKWORKS Inc., CFPM300, CFPM301, CFPM400, CFPM401, CFPMB301, CFUR301, CFUR401 and CFLU401 Imaging Tables can become lodged under the base or under the column of the table. |
| Root cause | Device Design |
| Action | OAKWORKS, Inc. issued a MEDICAL DEVICE ADVISORY NOTICE dated 8/18/2017 to customers providing more information in regards to the Medical Device Advisory Notice sent on May 12, 2017. The advisory notice was initiated in response to reported issues with the foot control. OakWorks, Inc. customer service will place the order for the appropriate Foot Control Kit listed below based on what table is impacted and the date of manufacture. PN 2015-204 - Foot Control Kit - Standard Riser 2015-209 - Foot Control Kit - Thinner Riser 2015-210 - Foot Control Kit - Holder Only 2015-211 - Foot Con |
| Quantity | 1,104 |
| Distribution | USA (nationwide) Distribution |
| FDA record | Open on fda.gov |