Recall Z-0628-2018 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Zimmer Biomet, Inc.

AGC 2000 LP Patella, labeled as AGC Porous Patellar Med, Model Number 150804. Product Usage: For use in Total Knee Arthroplasty.

DateAug 25, 2017
StatusTerminated
ReasonThe affected AGC 2000 LP Patella products were over-molded leaving the outside diameter nonconforming to the print dimension.
Root causeProcess change control
ActionConsignees were notified via mail on approximately 08/25/2017. Instructions to distributors, sales representatives, and distributor operations managers include to ensure affected team members are aware of the recall, locate and quarantine affected product in inventory, complete and return the Certification of Acknowledgement, return all affected product, and provide any additional hospitals and/or surgeons that received direct shipments. Instructions to risk managers and surgeons include to ensure affected personnel are informed of the recall, quarantine all affected product, return any aff
Quantity11 devices
DistributionUS Nationwide Distribution to IN, MO, NY, TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K833921AGC TOTAL KNEE PROSTHESIS
Biomet, Inc., Jan 31, 1984