Recall Z-0809-2018 ยท JXG Shunt, Central Nervous System And Components

Aesculap Implant Systems LLC

proSA Tool Set used to adjust and verify the proSA Adjustable Shunt System Shunt, central nervous system and components

DateSep 21, 2017
StatusTerminated
ReasonA proSA adjusting pin used to adjust the proSA Adjustable Shunt System does not set the correct valve opening pressure.
Root causeProcess control
ActionAesculap sent an Urgent Medical Device Recall Notification letter dated September 21, 2017. 1. Review the Urgent Medical Device Recall Notification Letter in its entirety and ensure that all users in your organization and other concerned persons are informed about this recall and immediately quarantine the unit. 2. Determine the current inventory of the affected lot(s) within your facility. Please quarantine the product and do use. Do not destroy any affected product. A Sales Representative will come to your facility and remove the affected sets and provide you replacements. 3. Please
Quantity2
DistributionUS Distribution to PA.