ClearanceBrief
Recall Z-0513-2018 ยท KPS System, Tomography, Computed, Emission

Philips Medical Systems (Cleveland) Inc

Precedence 6 Model # 882351 Product Usage: Precedence (cleared as Griffin) is an imaging system combining the acquisition of single photon nuclear medicine images and images from an x-ray computed tomography system. Precedence may produce non-attenuation corrected and attenuation corrected images of the distribution of radiopharmaceuticals in the body as well as x-ray transmission images. The CT

DateFeb 3, 2018 (initiated Dec 15, 2017)
StatusTerminated
ReasonDuring collimator exchange, the alignment of the collimator cart and detectors may cause a collimator exchange fail, and a collimator may drop to the floor beside the collimator cart. The collimator may be partially supported by the cart (e.g. leaning against the cart), or it may fall/slide all the way to the floor.
Root causeUnder Investigation by firm
ActionPhilips sent an Urgent Field Safety Notice letter dated December 15, 2017 to affected customers. The letter identified the affected product, problem and actions to be taken. For questions contact your local Philips representative.
Quantity1387 in total
DistributionWorldwide Distribution - US Nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041218GRIFFIN SPECT/CT IMAGING SYSTEM
Adac Laboratories, May 24, 2004