Recall Z-0558-2018 ยท NEY System, Ablation, Microwave And Accessories

Covidien LLC

Covidien Emprint Percutaneous Antenna with Thermosphere Technology, 20cm, Material CA20L1

DateAug 7, 2017
StatusOpen, Classified
ReasonThe firm received consumer reports of the ceramic trocar tip of the Emprint ablation antenna disengaging from the needle shaft post-ablation. This can result in the tip remaining in the patient.
Root causeDevice Design
ActionAll consignees were notified via Federal Express or certified mail on August 7, 2017, and the letter informs customers of the possible trocar tip disengagement and the actions they should take. Customers are requested to segregate and return affected product and acknowledge receipt and understanding of the Urgent Medical Device Recall Notice.
DistributionNationwide including PR, Canada, China
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133821EMPRINT ABLATION SYSTEM
Covidien, LLC, Apr 28, 2014