Recall Z-0661-2018 ยท MUE Full Field Digital, System, X-Ray, Mammographic
Fujifilm Medical Systems U.S.A., Inc.
ASPIRE Cristalle (for FDR-3000AWS) The Fujifilm Digital Mammography System, ASPIRE Cristalle (FDR MS-3500) generates full field digital mammography images that can, as other full-field digital mammography systems, be used for screening and diagnosis of breast cancer and is intended for use in the same clinical applications as traditional screen-film mammography systems.
| Date | Aug 4, 2017 |
| Status | Terminated |
| Reason | FUJIFILM Medical Systems U.S.A., Inc. (FMSU) identified a potential failure with our Mammography system. The acquisition workstations, FDR1000AWS, FDR-2000AWS, FDR-3000AWS, and CRIR363AWS assign a unique ID number to every image study but very rarely, with the acquisition workstation software versions, VS.O, VS.1, VS.2 , V6.0, V6.1, and V7.0, the system may assign the same ID number to a new set of images that was already assigned to the previous set of images due to the error in ID number generation logic. If an Image with this error (with duplicate ID number) Is transmitted to PACS, it may o |
| Root cause | Software design |
| Action | The firm, FUJIFILM Medical Systems U.S.A., Inc. (FMSU), sent an "URGENT: MEDICAL DEVICE CORRECTION" letter dated 7/28/2017 to its customers. The letter described the products, problem and actions to be taken. The customers were instructed to please read and follow the instructions in the "ACTIONS TO BE TAKEN BY CUSTOMER/USER" section. If you have an error, contact your local FUJIFILM office; and complete and return the Customer Feedback Form. FUJIFILM service personnel will contact all of the medical facilities where the applicable products have been installed to arrange for this correction a |
| Quantity | 71 units (68 units Domestic. 3 units Foreign) in total |
| Distribution | Worldwide Distribution. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K133972 | ASPIRE CRISTALLE Fujifilm Medical Systems U.S.A, Inc., Mar 25, 2014 |