ClearanceBrief
Recall Z-1345-2018 ยท FMG Stopcock, I.V. Set

ICU Medical, Inc.

7" Smallbore Ext Set w/MicroClave, Clave T-Connector, Clamp, Rotating Luer, 50 pouches per case, Item B9900-629, Sterile, Rx. The firm name on the label is icumedical, San Clemente, CA. The Intravascular Administration Set is a single use, sterile, non-pyrogenic device which provides access for the administration of fluids from a container to a patient s vascular system through the administrat

DateFeb 5, 2018
StatusTerminated
ReasonThe infusion sets were manufactured with an incorrect spin collar.
Root causeProcess control
ActionThe firm, ICU Medical Inc., sent an "URGENT: Medical Device Recall Notification" letter dated 2/5/2018 to its customers via UPS on 2/5/2018 in the U.S. The letter described the product, problem and actions to be taken. The customers were instructed to: inspect your inventory for specific product and lot numbers; quarantine all affected devices; complete and return the Recall Response Form as applicable (even if you have no affected devices) via Fax to: 1-866-899-7474 or email to: ICUMedical2805@stericycle.com; if you have distributed the product further, notify your accounts that received the
Quantity2,950 pouches
DistributionWorldwide Distribution: US (nationwide) to state in AZ, CA, CO, FL, SC, and TX; and countries of: Canada and Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K964435PRIMARY IV SET
Icu Medical, Inc., Apr 2, 1997