Recall Z-0969-2018 ยท JJY Multi-Analyte Controls, All Kinds (Assayed)

Randox Laboratories

Immunoassay Premium Controls (3x4x5mL and 12x5mL packages)

DateAug 11, 2017
StatusTerminated
ReasonThe ACTH in Immunoassay Premium and Premium Plus quality control does not meet the quoted reconstituted stability claim in the product IFU.
Root causeDevice Design
ActionOn July 24, 2017 an Urgent notice was issued instructing customers to place a copy a notice in all kits remaining in stock. This notifies customers that they should amend the claims for open stability in the instructions for use, to state use immediately after the 30 minute reconstitution period. Questions or concerns can be emailed to technical.services@randox.com
Quantity117 Kits
DistributionDistribution US nationwide.