Recall Z-1269-2018 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Roche Diagnostics Corporation

Modular E 170 Analyzer: Immunoassay Chemistry Analyzer, Chemistry (Photometric, Discrete) for clinical use, Catalog Numbers: 03739040692, 03023109001, 05023599001, 05023572001, 03739040001 & 03023109973.

DateDec 18, 2017
StatusTerminated
ReasonComplaints have been received concerning questionable results on analyzers due to customers utilizing sample 13 mm tubes without tube adapters.
Root causeLabeling False and Misleading
ActionRoche Diagnostics Corporation issued Urgent Medical Device Correction (UMDC) notices and Fax Forms to their customers dated 12/18/2017 via UPS Ground (receipt required). Customers are encouraged to do the following Mandatory Steps to improve the vertical alignment and handling of 13 mm sample tubes and false bottom tubes: FOR cobas e 411 & cobas e 601 analyzers and MODULAR ANALYTICS E 170 module: 1) Always use Roche rack cup adapters with 13 mm tubes 2) Ensure sample tubes are straight FOR cobas e 411 disk system: 1) Always use 13 mm Sample Disk Tube Adapters (SDTAs) with 13 mm sample tube
DistributionUS Nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K953239BOEHRINGER MANNHEIM/HITACHI 917 ANALYZER
Boehringer Mannheim Corp., Sep 25, 1995