ClearanceBrief
Recall Z-0814-2018 ยท ISY Joint, Knee, External Limb Component

Otto Bock Healthcare Product

OTTO BOCK Kenevo knee joints Model 3C60

DateSep 12, 2017
StatusTerminated
ReasonOtto Bock Healthcare Products GmbH has identified a design issue through extended durability testing where the pylon clamping mechanism has the potential to cause damage to the pylon. When used longer than 1.45 million cycles (which represents approximately 2.3 years of use of the knee joint under heavy usage conditions), AND the pylon clamp is overtightened according to the labeled torque specification, AND the patient weighs at most 125kg it may be remote that the pylon breaks, and causing the patient to fall. There have been no failures or complaints reported directly attributed to the fail
Root causeDevice Design
ActionOn September 14, 2017 a letter was sent to all their consignees via 2nd day signature receipt required to customers instructing them to contact their patients and send in their devices for service within the first 2.3 years of use. The pylon clamp will be replaced at no charge. The pylon will also be inspected and replaced if necessary. The customer will be requested to complete a reply form indicating they have received the notice and intend to contact their patients to arrange for the requested service. Customer Service: (800) 328-3458 Option 1 Loaner Service: (800) 328-3458 Option 3, then
Quantity75 units
DistributionAK, AR, AZ, CA, CT, DC, GA, IL, IN, KY, LA, MD, MI, MN, NC, ND, NJ, MN, NY, OH, OK, OR, PA, PR, TX, VA, and WA AU, BE, DE, FR, GB, IT, KW, LU, NL, NO, AT, SA, SE, CH, ES, ZA, and HU
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.