ClearanceBrief
Recall Z-0722-2018 ยท LPH Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Porous Uncemented

Zimmer Biomet, Inc.

Versys 6 inch Beaded Full Coat Plus Hip Prosthesis. Labeled as follows: a. VERSYS 6 INCH BEADED FC 19X160MM STD BODY STD NECK (Item No. 00894301906); b. VERSYS 6 INCH BEADED FC 21X160MM STD BODY STD NECK (Item No. 00784302106); c. VERSYS 6 INCH BEADED FC 22X160MM STD BODY STD NECK (Item No. 00784302206); d. VERSYS 6 INCH BEADED FC STEM 19X160MM LM (Item No. 00784301936); e. VERSYS 6 INC

DateAug 1, 2017
StatusTerminated
ReasonThe products being recalled exceeded the weight previously tested leading to a possibility of damage to the package and/or a compromised sterile barrier.
Root causePackage design/selection
ActionThe firm, Zimmer Biomet, sent an "URGENT MEDICAL DEVICE RECALL REMOVAL"-Lot Specific letter dated 8/1/2017 to its Customers via email and/or FedEx. The letter describes the product, problem and actions to be taken. The Instructions to Distributors, Sales Representatives, and Distributor Operation Managers included to ensure affected team member are aware of the contents of the notification letter, immediately locate and quarantine affected product in inventory, complete and return the Certification of Acknowledgement form by email to: corporatequality.postmarket@zimmerbiomet.com., return all
Quantity38,069 units total
DistributionWorldwide Distribution: US (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, and Puerto Rico; and to countries of: Austria,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K953337BETA HIP PROSTHESIS
Zimmer, Inc., Jan 22, 1996
K973714VERSYS HIP SYSTEM BEADED HIP PROSTHESIS
Zimmer, Inc., Dec 24, 1997