ClearanceBrief
Recall Z-1316-2018 ยท ILQ System, Communication, Powered

Lifeline Systems Company

HomeSafe AutoAlert Pendant works in conjunction with a compatible Lifeline communicator.

DateNov 13, 2017
StatusTerminated
ReasonA programing error in some Model FD100 HomeSafe AutoAlert Pendants will render the fall detection feature inoperable.
Root causeSoftware Manufacturing/Software Deployment
ActionFirm has contacted all recipients and informed them of an issue that affects Philips Lifeline AutoAlert help button from automatically signaling the Lifeline Response Center for help in the event of a fall. Further instructed them that they would need to press the Lifeline Help Button in order to connect with the emergency response center. They then instruct them to return their inventory for replacement. New AAHB Replacements will be provided when available. Philips will work with the Programs to ensure buttons swaps are completed.
Quantity20201
DistributionUS and Canada distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K914103LIFELINE(R) PERSONAL RESPONSE SYSTEM
Lifeline Systems, Inc., Sep 27, 1991