ClearanceBrief
Recall Z-1169-2018 ยท JQP Calculator/Data Processing Module, For Clinical Use

Beckman Coulter Inc.

AutoMate 1200 Catalog Number ODL25120; AutoMate 1250 Catalog Number ODL25125; AutoMate 2500 Catalog Number ODL25250; AutoMate 2550 Catalog Number ODL25255

DateJan 19, 2018
StatusCompleted
ReasonBelts on the x-axis (and on the y-axis, to a lesser extent) manufactured from 2015 are showing premature failure, posing the remote risk of cross-contamination.
Root causeDevice Design
ActionBeckman Coulter plans to replace the belt via Service Modification during the next planned service maintenance visit. At that time, the customer will be informed of the belt replacement by the service engineer.
Quantity98 units in the US
DistributionAustralia, Austria, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Canada, China, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Gibraltar, Greece, Hong Kong, Hungary, India, Ireland, Israel, Italy, Korea, Kuwait, Latvia, Lithuania, Macao, Mayotte, Mexico, Netherlands, New Ze
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.